Objective analysis of contact lens fit

Lurdes Belda-Salmerón, Tom Drew, Lee Hall, James Stuart Wolffsohn*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose: To assess the validity and repeatability of objective compared to subjective contact lens fit analysis.

Methods: Thirty-five subjects (aged 22.0. ±. 3.0 years) wore two different soft contact lens designs. Four lens fit variables: centration, horizontal lag, post-blink movement in up-gaze and push-up recovery speed were assessed subjectively (four observers) and objectively from slit-lamp biomicroscopy captured images and video. The analysis was repeated a week later.

Results: The average of the four experienced observers was compared to objective measures, but centration, movement on blink, lag and push-up recovery speed all varied significantly between them (p <. 0.001). Horizontal lens centration was on average close to central as assessed both objectively and subjectively (p > 0.05). The 95% confidence interval of subjective repeatability was better than objective assessment (±0.128. mm versus ±0.168. mm, p = 0.417), but utilised only 78% of the objective range. Vertical centration assessed objectively showed a slight inferior decentration (0.371. ±. 0.381. mm) with good inter- and intrasession repeatability (p > 0.05). Movement-on-blink was lower estimated subjectively than measured objectively (0.269. ±. 0.179. mm versus 0.352. ±. 0.355. mm; p = 0.035), but had better repeatability (±0.124. mm versus ±0.314. mm 95% confidence interval) unless correcting for the smaller range (47%). Horizontal lag was lower estimated subjectively (0.562. ±. 0.259. mm) than measured objectively (0.708. ±. 0.374. mm, p <. 0.001), had poorer repeatability (±0.132. mm versus ±0.089. mm 95% confidence interval) and had a smaller range (63%). Subjective categorisation of push-up speed of recovery showed reasonable differentiation relative to objective measurement (p <. 0.001).

Conclusions: The objective image analysis allows an accurate, reliable and repeatable assessment of soft contact lens fit characteristics, being a useful tool for research and optimisation of lens fit in clinical practice.

Original languageEnglish
Pages (from-to)163-167
Number of pages5
JournalContact Lens and Anterior Eye
Volume38
Issue number3
Early online date26 Feb 2015
DOIs
Publication statusPublished - 2015

Bibliographical note

© 2015, Elsevier. Licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International http://creativecommons.org/licenses/by-nc-nd/4.0/

Funding: Ministerio de Educación, Cultura y Deporte research scholarship to Lurdes Belda-Salmerón (FPU#AP2010-2107#)

Keywords

  • fitting characteristics
  • lens evaluation
  • objective repeatability
  • soft contact lens
  • validity

Fingerprint

Dive into the research topics of 'Objective analysis of contact lens fit'. Together they form a unique fingerprint.

Cite this