A Cadaveric Study Validating in vitro Monitoring Techniques to Measure the Failure Mechanism of Glenoid Implants against Clinical CT

Sarah Junaid, Thomas Gregory, Shirley Fetherston, Roger Emery, Andrew Amis, Ulrich Hansen

Research output: Contribution to journalArticlepeer-review

Abstract

Definite glenoid implant loosening is identifiable on radiographs, however, identifying early loosening still eludes clinicians. Methods to monitor glenoid loosening in vitro have not been validated to clinical imaging. This study investigates the correlation between in vitro measures and CT images. Ten cadaveric scapulae were implanted with a pegged glenoid implant and fatigue tested to failure. Each scapulae were cyclically loaded superiorly and CT scanned every 20,000 cycles until failure to monitor progressive radiolucent lines. The superior and inferior rim displacements were also measured. A finite element (FE) model of one scapula was used to analyse the interfacial stresses at the implant/cement and cement/bone. All ten implants failed inferiorly at the implant-cement interface, two also failed at the cement-bone interface inferiorly, and three showed superior failure. Failure occurred at of 80,966 ± 53,729 (mean ± SD) cycles. CT scans confirmed failure of the fixation, and in most cases, was observed either before or with visual failure, indicating its capacity to detect loosening earlier for earlier intervention if needed. Significant correlations were found between both increasing inferior rim displacement (ASTM standard F2028-14), increasing vertical head displacement and failure of the glenoid implant. The FE model showed peak tensile stresses inferiorly and high compressive stresses superiorly, corroborating experimental findings. Similar failure modes have been cited in clinical and in vitro studies. In vitro monitoring methods correlated to failure progression in clinical CT images.
Clinical Significance: The study highlights failure at the implant-cement interface and early signs of failure are identifiable in CT images.
Original languageEnglish
Pages (from-to)2524-2532
JournalJournal of Orthopaedic Research
Volume36
Issue number9
Early online date25 Apr 2018
DOIs
Publication statusPublished - 1 Sept 2018

Bibliographical note

© 2018 The Authors. Journal of Orthopaedic Research ® Published by Wiley Periodicals, Inc. on behalf of the Orthopaedic Research Society.

This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited

Funding: Arthritis Research UK.

Keywords

  • glenoid loosening, fixation failure, CT, radiolucent lines

Fingerprint

Dive into the research topics of 'A Cadaveric Study Validating in vitro Monitoring Techniques to Measure the Failure Mechanism of Glenoid Implants against Clinical CT'. Together they form a unique fingerprint.

Cite this