T1 - Repeatability of ocular aberration measurements in patients with keratoconus
AU - Jinabhai, Amit
AU - Radhakrishnan, Hema
AU - O'Donnell, Clare
PY - 2011/11
Y1 - 2011/11
N2 - Purpose: To explore the repeatability of lower-order and higher-order ocular aberrations measured in patients with keratoconus. Methods: The IRX-3 (Imagine Eyes, Paris, France) aberrometer was used to record lower-order and higher-order aberrations in 31 eyes of 31 patients with keratoconus. Four monocular measurements were taken consecutively for each patient. The aberrometry data were analysed up to the 5th Zernike order for a 4-mm pupil diameter. The data were evaluated using repeated-measures anova and Friedman analyses. Repeatability was analysed using within-subject standard deviation (SW) and the repeatability limit (r) calculated as 1.96 ×√2×Sw. Results: Of the 11 aberration terms evaluated, the repeatability of Z (2,0) (mean= 1.36μm; SW=0.09μm; r=0.26μm); Z (2,±2) RMS (mean=1.05μm; SW= 0.09μm; r=0.24μm) and Z (4,0) aberrations (mean=0.34μm; SW=0.09 μm; r=0.24μm) showed the highest variability. In contrast, Z (3,±1) RMS aberrations (mean=0.85μm; SW=0.06μm; r=0.16μm) and Z (4,±2) RMS aberrations (mean=0.40μm; SW=0.07μm; r=0.18μm) showed comparatively better repeatability. Conclusions: The lower-order and higher-order aberrations measured in this group of keratoconic patients showed higher levels of variability compared to previous investigations of visually-normal subjects. These results may be of interest to eyecare practitioners involved in the design and fitting of aberration-controlling contact lenses for patients with keratoconus.
AB - Purpose: To explore the repeatability of lower-order and higher-order ocular aberrations measured in patients with keratoconus. Methods: The IRX-3 (Imagine Eyes, Paris, France) aberrometer was used to record lower-order and higher-order aberrations in 31 eyes of 31 patients with keratoconus. Four monocular measurements were taken consecutively for each patient. The aberrometry data were analysed up to the 5th Zernike order for a 4-mm pupil diameter. The data were evaluated using repeated-measures anova and Friedman analyses. Repeatability was analysed using within-subject standard deviation (SW) and the repeatability limit (r) calculated as 1.96 ×√2×Sw. Results: Of the 11 aberration terms evaluated, the repeatability of Z (2,0) (mean= 1.36μm; SW=0.09μm; r=0.26μm); Z (2,±2) RMS (mean=1.05μm; SW= 0.09μm; r=0.24μm) and Z (4,0) aberrations (mean=0.34μm; SW=0.09 μm; r=0.24μm) showed the highest variability. In contrast, Z (3,±1) RMS aberrations (mean=0.85μm; SW=0.06μm; r=0.16μm) and Z (4,±2) RMS aberrations (mean=0.40μm; SW=0.07μm; r=0.18μm) showed comparatively better repeatability. Conclusions: The lower-order and higher-order aberrations measured in this group of keratoconic patients showed higher levels of variability compared to previous investigations of visually-normal subjects. These results may be of interest to eyecare practitioners involved in the design and fitting of aberration-controlling contact lenses for patients with keratoconus.
KW - aberrations
KW - coma RMS
KW - higher-order
KW - keratoconus
KW - lower-order
KW - repeatability
UR - http://www.scopus.com/inward/record.url?scp=80053653269&partnerID=8YFLogxK
UR - http://onlinelibrary.wiley.com/doi/10.1111/j.1475-1313.2011.00868.x/abstract
U2 - 10.1111/j.1475-1313.2011.00868.x
DO - 10.1111/j.1475-1313.2011.00868.x
M3 - Article
C2 - 21936862
AN - SCOPUS:80053653269
SN - 0275-5408
VL - 31
SP - 588
EP - 594
JO - Ophthalmic and Physiological Optics
JF - Ophthalmic and Physiological Optics
IS - 6
ER -